| Model Number |
71674074 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the 3.2mm t-handle trocar confirmed the handle of the device is broken off.The broken piece was not returned with the device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the pro failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an internal fixation surgery, the 3.2mm t-handle trocar broke.Incident occurred while instrument was outside the patient and no pieces fell into the incision.Surgery was resumed, without any delay, using a back-up device.Patient was not injured as consequence of this problem.
|
| |
|
Search Alerts/Recalls
|
|
|