MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Low Blood Pressure/ Hypotension (1914); Perforation (2001)
Date of Event 10/19/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a perforation occurred in the right external iliac.A 9mm x 40mm epic stent, a 7mm x 60mm eluvia stent and 7mm x 40mm mustang were selected for the treatment of peripheral vascular disease in the external iliac.The lesion was pre-dilated with a 5mm x 100mm mustang balloon followed by an arteriogram to mark the landing zones.A 9mm x 40mm epic stent and a 7mm x 60mm eluvia stent were placed to treat the lesion.Post dilation was performed with a 7mm x 40mm mustang.The certified rn anesthetist noticed changes in the patient's carbon dioxide levels and blood pressure.Repeat arteriograms were performed reveling a perforation in the right external iliac.Hypotension and blood loss were reported as a result of the perforation.It was thought that the perforation possibly occurred during post dilation of the stents and calcification of the lesion contributed to the perforation.The perforation was treated using a non-boston scientific stent and multiple long balloon inflations.There were no patient complications reported after the perforation was treated.The patient was stable after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12802589
Report Number2134265-2021-13997
Device Sequence Number5484596
Product Code NIO
UDI-Device Identifier08714729805793
UDI-Public08714729805793
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/18/2026
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0027340789
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/19/2021
Initial Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexFemale
Patient RaceBlack Or African American
-
-