| Model Number |
39200 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Low Blood Pressure/ Hypotension (1914); Perforation (2001)
|
| Date of Event |
10/19/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a perforation occurred in the right external iliac.A 9mm x 40mm epic stent, a 7mm x 60mm eluvia stent and 7mm x 40mm mustang were selected for the treatment of peripheral vascular disease in the external iliac.The lesion was pre-dilated with a 5mm x 100mm mustang balloon followed by an arteriogram to mark the landing zones.A 9mm x 40mm epic stent and a 7mm x 60mm eluvia stent were placed to treat the lesion.Post dilation was performed with a 7mm x 40mm mustang.The certified rn anesthetist noticed changes in the patient's carbon dioxide levels and blood pressure.Repeat arteriograms were performed reveling a perforation in the right external iliac.Hypotension and blood loss were reported as a result of the perforation.It was thought that the perforation possibly occurred during post dilation of the stents and calcification of the lesion contributed to the perforation.The perforation was treated using a non-boston scientific stent and multiple long balloon inflations.There were no patient complications reported after the perforation was treated.The patient was stable after the procedure.
|
| |
|