• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. ETHLN BLK 18IN 4-0 S/A P-3 PRM MP; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHLN BLK 18IN 4-0 S/A P-3 PRM MP; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Model Number 699H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient underwent a neck procedure on (b)(6) 2021 and the suture was used.During skin closure, the needle popped off the suture.Another suture was used to complete the procedure.There were no adverse consequences for the patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).A manufacturing record evaluation was performed for the finished device lot number pcb924 / 699h, and no non-conformance's related to the reported complaint condition were identified.(b)(4).An opened box with thirty-three packets of product code 699h was returned for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packet.In order to evaluate the conditions of thirteen returned samples, the packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed, and the pull force meets the requirements.Events were submitted via 2210968-2021-11350.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 18IN 4-0 S/A P-3 PRM MP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
982 road 183 km 8.3
san lorenzo PR
*  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key12803716
Report Number2210968-2021-11351
Device Sequence Number20008297
Product Code GAR
UDI-Device Identifier10705031008632
UDI-Public10705031008632
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model Number699H
Device Catalogue Number699H
Device Lot NumberPCB924
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2021
Initial Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-