|
It was reported that, during a thr surgery, the femoral head impactor ((b)(6)) broke on impaction.All pieces were retrieved.Surgery was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.
|
|
H3, h6: the device was returned for evaluation.A visual inspection of the returned device confirms the tip fractured into two pieces.Both pieces were returned for evaluation.The device shows signs of significant wear and use.The manufacturing process of the device did not contribute to the reported event.A medical investigation was conducted and confirms per case details, all pieces were retrieved from the patient.The procedure was completed using a backup device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this product and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|