U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. ALLEVYN LIFE L 15.4X15.4 CTN10; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. ALLEVYN LIFE L 15.4X15.4 CTN10; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 66801069
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code Skin Disorders (4543)
Date of Event 10/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case- (b)(4).
 
Event or Problem Description
It was reported that during a sarcoma resection surgery an allevyn life l 15.4x15.4 ctn10 was used on the patient, the silicone plates was placed in the bony prominences accordantly to the institutional protocol patient was in ventral position was required during procedure, surgery lasted for 1 hour and 30 minutes and when returning patient to ddh position and after removing the plates it was noticed that the patient's skin had macerations and looked like a kind of "glue".This problem was treated using 5 sachets of sting free adhesive remover.
 
Additional Manufacturer Narrative
Additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The new information states that the wound was not macerated and just had silicone residues.Therefore, it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN LIFE L 15.4X15.4 CTN10
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12804752
Report Number8043484-2021-01953
Device Sequence Number4778017
Product Code NAD
UDI-Device Identifier05000223478049
UDI-Public05000223478049
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/01/2023
Device Model Number66801069
Device Catalogue Number66801069
Device Lot Number202029
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2021
Supplement Date Received by Manufacturer12/01/2021
Initial Report FDA Received Date11/12/2021
Supplement Report FDA Received Date12/08/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-