BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that tip detachment occurred.A 9x40x75 epic vascular stent was selected for use.However, during preparation, it was noted that the tip of the device broke off before loading onto the wire.The stent was never inserted into patient.The procedure was completed wit another epic stent of the same size.No patient complications were reported.
 
Event or Problem Description
It was reported that tip detachment occurred.A 9x40x75 epic vascular stent was selected for use.However, during preparation, it was noted that the tip of the device broke off before loading onto the wire.The stent was never inserted into patient.The procedure was completed wit another epic stent of the same size.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of an epic self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the stent is partially deployed 5mm from the sheath.The inner liner is kinked 19mm and 25mm from the tip.Microscopic examination revealed no additional damages.Because there was no evidence of any product quality deficiencies, it was considered likely that the kinks and partial deploy were attributable to failure to follow instructions.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12804902
Report Number2134265-2021-13861
Device Sequence Number17529780
Product Code NIO
UDI-Device Identifier08714729805793
UDI-Public08714729805793
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2026
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0027680092
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/27/2021
Supplement Date Received by Manufacturer11/18/2021
Initial Report FDA Received Date11/12/2021
Supplement Report FDA Received Date11/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-