• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 QUICKSET ACE GRATER HEAD 70MM; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 QUICKSET ACE GRATER HEAD 70MM; HIP INSTRUMENTS : REAMERS Back to Search Results
Model Number 2440-00-570
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the graters are dull by the sales rep.Upon follow-up, only two grater heads within the set were identified as dull within the healthcare setting by the surgeon.The remaining graters were returned for routine product replenishment only and a failure was not identified at the time.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was received for examination.Physical evaluation of the returned instrument was able to identify dullness of the device.Cutting surface is dull to the touch, is not as sharp as first use condition.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HEAD 70MM
Common Device Name
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12805365
Report Number1818910-2021-25005
Device Sequence Number4698946
Product Code HTO
UDI-Device Identifier10603295124177
UDI-Public10603295124177
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2440-00-570
Device Catalogue Number244000570
Device Lot NumberSO2047971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2021
Initial Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/11/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-