• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PICOSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PICOSURE Back to Search Results
Model Number 100-7012-010
Medical Device Problem Code Radiation Underexposure (3018)
Health Effect - Clinical Code Blister (4537)
Date of Event 06/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a second patient experienced blisters following a pigmentation treatment using picosure.First patient was reported in mdr# 1222993-2021-00043.
 
Additional Manufacturer Narrative
Blisters are an expected transient side effect.No medical intervention was reported.Blisters has resolved.During a device evaluation, technician confirmed that the energy was low and greater than 20% outside of specification.The device history record confirms that the product passed and met all requirements before releasing.Technician repaired the device.Since the energy was confirmed to be outside of the specification, this event is an aro (accidental radiation occurrence) and therefore reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICOSURE
Common Device Name
PICOSURE
Manufacturer Contact
5 carlisle road
westford, MA 01886
MDR Report Key12805627
Report Number1222993-2021-00044
Device Sequence Number14618049
Product Code GEX
UDI-Device Identifier00841494100001
UDI-Public(01)00841494100001(11)171212
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K210226
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100-7012-010
Device Catalogue Number105-7012-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2021
Initial Report FDA Received Date11/12/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-