| Model Number |
720074-03 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a surgical procedure to explant this spectra penile prosthesis (spp) as it was said to be broken.The device was removed and replaced with an inflatable penile prosthesis (ipp).There were no patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: based on the information available, the cause that contributed to the reported broken device cannot be established as the product is not available for analysis.Device history record (dhr): the device history record (dhr) confirmed that the device met all material, assembly and performance specifications.Device technical analysis: the device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Labeling review: a review of the spp instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.Investigation conclusion: based on the information available, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a surgical procedure to explant this spectra penile prosthesis (spp) as it was said to be broken.The device was removed and replaced with an inflatable penile prosthesis (ipp).There were no patient complications.
|
| |
|