PENUMBRA, INC. PENUMBRA ENGINE; QEW Back to Search Results
Model Number PMXENGN-A
Medical Device Problem Code Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2021-02575.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the iliac vein and inferior vena cava (ivc) using a penumbra engine (engine) and a lightning aspiration tubing (lightning).During the procedure, while aspirating clot using the engine and the lightning, the engine lost power towards the end of the procedure.To troubleshoot, the engine was power cycled, and the lightning was unplugged and re-plugged into the engine; however, the issue was not resolved.It was reported that a new lightning was plugged into the engine; however, the lightning unit did not power on.Therefore, the first lightning and engine were not used for the remainder of the procedure.Next, the physician plugged the new lightning into another engine and the lightning unit powered on; however, upon powering on the engine, it was reported only two of the four indicator lights illuminated, and the indicator lights were very dim.The procedure was completed using the new lightning and the second engine with only two of the four indicator lights illuminated.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the device associated with this complaint was expected to be returned; however, additional information received from the penumbra sales representative indicated that the device was inadvertently discarded and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-02575.H3 other text: placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA ENGINE
Common Device Name
QEW
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key12807240
Report Number3005168196-2021-02574
Device Sequence Number5485238
Product Code QEW
UDI-Device Identifier00815948020023
UDI-Public00815948020023
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K180105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/07/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMXENGN-A
Device Catalogue NumberPMXENGN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2021
Supplement Date Received by Manufacturer12/07/2021
Initial Report FDA Received Date11/12/2021
Supplement Report FDA Received Date12/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/27/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age44 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-