| Model Number |
9004051 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the comport and signals are incorrect on a smiths medical capnograph.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol b10010116, as a result of warning letter cms# (b)(4).No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.There is no physical damage on the device.The technician performed an analog output function test.The reported complaint was verified after test the analog output function test on the returned 9004051 monitor and was not able to get 1 volts out of channel 1 and 2.Due to the failure of the analog output test the main board will need to be replace.The root cause of the reported issue was found to be a defective main board.The technician replaced main board.
|
| |
|
Event or Problem Description
|
|
Additional information received via email 05-nov-2021: discovered while in use.
|
| |
|