| Model Number |
AVSM08060 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Restenosis (4576)
|
| Date of Event |
09/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 12/2022).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical trial, that approximately three months and twenty two days post index procedure, the subject developed 25% of target lesion stenosis.Standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation and image was not provided.Based on the information available the investigation is closed with inconclusive result.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risk.Based on the instructions for use complications and adverse events associated with the use of the covera vascular covered stent may include the usual complications associated with endovascular stent and covered stent placement and dialysis shunt revisions, which includes restenosis of the target vessel.Regarding pre and post dilation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated" and "post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.Avoid balloon dilation in the healthy, non-stenosed segment of the vessel." h10: d4 (expiry date: 12/2022).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported through the results of a clinical trial, that approximately three months and twenty two days post index procedure, the subject developed 25% of target lesion stenosis.A standard percutaneous transluminal angioplasty (pta) was used to successfully treat the target lesion.The current patient status was not provided.
|
| |
|
Search Alerts/Recalls
|