| Model Number |
1053958 |
| Medical Device Problem Code |
Inadequate Lubrication (4057)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that they have experienced that the vantage arc grib on the right side seems compromised.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The user notified elekta that they had noticed that the ring locking mechanism had been compromised on the right side on several occasions.No movement during treatment has been reported.No patient injury or effect on implantation procedure had been seen.The support arc and the z-scales were sent to elekta for investigation.The investigation found that when mounting the parts together it was clear that the system has not been properly lubricated in combination with a worn system and not replacing the worn knobs.The locking was substantially improved and was working correctly after lubrication of the support arc according to the instructions for use.The support arc has been replaced and the user has been thoroughly informed regarding the importance of lubrication according to the instruction for use.The product works as designed and intended when standard maintenance is applied.The root cause was determined to be use error.
|
| |
|