• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED POSTERIOR STABILIZED (PS) NARROW LEFT SIZE 6; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED POSTERIOR STABILIZED (PS) NARROW LEFT SIZE 6; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Damage to Ligament(s) (1952); Pain (1994); Swelling/ Edema (4577)
Date of Event 10/05/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Initial left total knee arthroplasty approximately 4 years ago, mua on one month later due to stiffness, 1st revision approximately 3 years later due to instability and subsequently 2nd revision last month due to pain and swelling after a fall.The patient fell due to unknown reasons.The tibia and femur components were exchanged without complications, patella remained intact.
 
Additional Manufacturer Narrative
(b)(4).Item#: 42532006701; tibia cemented 5 degree stemmed left size d; lot#: 63727290.Item#: 42512600514; articular surface fixed bearing (cps) left 14mm; customer has indicated that the product will not be returned to zimmer biomet for investigation, as it has been discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2021-00288 , 3007963827-2021-00287.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6.Reported event was unable to be confirmed due to limited information received from the customer.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device history record (dhr) was reviewed and no discrepancies were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: possible periprosthetic lucency surrounding the femoral component; also possible periprosthetic lucency surrounding the anterior of the tibial component; image 1 demonstrates soft tissue swelling; otherwise unremarkable appearance; overall fit and alignment appears normal; no wear or contributing factors; no causes for pain are identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED POSTERIOR STABILIZED (PS) NARROW LEFT SIZE 6
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12810667
Report Number3007963827-2021-00289
Device Sequence Number17639929
Product Code JWH
UDI-Device Identifier00889024228733
UDI-Public(01)00889024228733(17)270531(10)63690665
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42500006001
Device Lot Number63690665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2021
Supplement Date Received by Manufacturer11/17/2021
Initial Report FDA Received Date11/15/2021
Supplement Report FDA Received Date11/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/22/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight68 KG
-
-