Event or Problem Description
|
|
Surgeon could not get poly to seat.Claimed something was wrong with locking mechanism.There was a surgical delay of 5 minutes.Doi: unknown; dor: (b)(6) 2021; unknown side.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The photograph was reviewed, and no evidence of anything indicative of a device nonconformance was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|