U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. SELF-STARTING, VARIABLE SCREW; SIZE Ø4.5X16 MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. SELF-STARTING, VARIABLE SCREW; SIZE Ø4.5X16 MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 8801-04516DA
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while inserting an ozark self-starting variable screw into an ozark plate, a sliver piece of metal came out through the plate window as the screw was driven deeper.The procedure was completed successfully without surgical delay.No adverse consequences were reported.This record captures the ozark self-starting variable screw.
 
Additional Manufacturer Narrative
A visual inspection was performed which confirmed the ozark screw did fracture.The threading was returned as 2 fragments.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.A review of complaint history associated with the subject catalog number was performed, no similar complaints were identified.Due to screw main body and associated plate not being able to be returned, a definitive root cause could not be determined.Based on screw thread fragments returned, the most potential root cause may be screw over angulation, however this cannot be confirmed.Screw over angulation may cause screw thread to contact plate during insertion and strip the threads.
 
Event or Problem Description
It was reported that while inserting an ozark self-starting variable screw into an ozark plate, a sliver piece of metal came out through the plate window as the screw was driven deeper.The procedure was completed successfully without surgical delay.No adverse consequences were reported.This record captures the ozark self-starting variable screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF-STARTING, VARIABLE SCREW; SIZE Ø4.5X16 MM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key12812679
Report Number3004774118-2021-00357
Device Sequence Number20220837
Product Code KWQ
UDI-Device Identifier10888857393721
UDI-Public10888857393721
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K172104
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8801-04516DA
Device Catalogue Number8801-04516DA
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/19/2021
Supplement Date Received by Manufacturer01/26/2022
Initial Report FDA Received Date11/15/2021
Supplement Report FDA Received Date02/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-