| Model Number |
39200 |
| Medical Device Problem Codes |
Break (1069); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/02/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that shaft break and stent deployment issue occurred.The target lesion was located in the mildly tortuous and moderately calcified left iliac artery.Two 10x80x75 epic stents were used for treatment.However, during the procedure on both devices, the shaft came apart from the stent before fully deployment.Additional stent was deployed to make sure original stents stayed deployed and fully opened.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B5.Describe event or problem- corrected and updated.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break and stent deployment issue occurred.The target lesion was located in the mildly tortuous and moderately calcified left iliac artery.Two 10x80x75 epic stents were used for treatment.However, during the procedure on both devices, the shaft came apart from the stent before fully deployment.Additional stent was deployed to make sure original stents stayed deployed and fully opened.There were no patient complications nor injuries reported.It was further reported that just one 10x80x75 epic stent was used and not two epic stents as previously reported.There was resistance during advancing of the device and the shaft break occurred prior to stent deployment.
|
| |
|