• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST W/ ECHO; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST W/ ECHO; SURGICAL MESH Back to Search Results
Catalog Number 5955810
Medical Device Problem Codes Defective Device (2588); Patient Device Interaction Problem (4001); Migration (4003)
Health Effect - Clinical Codes Adhesion(s) (1695); Granuloma (1876); Pain (1994); Hernia (2240); Obstruction/Occlusion (2422); Insufficient Information (4580)
Date of Event 02/01/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient"; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventralight st mesh on (b)(6) 2016.As reported, the patient is making a claim for an adverse patient outcome against ventralight st mesh.Attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient." it is also alleged that the device was defective.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient"; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-jan-2015) is considered to be a best estimate.Updated fields: a2, b2, b3, b4, b5, b6, b7, d3, d4 (product catalog no., udi no., expiry date, corporate lot no.), d.6b (date of explant), g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventralight st mesh on (b)(6) 2016.As reported, the patient is making a claim for an adverse patient outcome against ventralight st mesh.Attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient." it is also alleged that the device was defective.Addendum per additional information provided: (b)(6) 2016 - patient was diagnosed with ventral hernia thereby underwent laparoscopic repair with implant of ventralight st using echo ps (device #1).Per op notes, ¿extensive lysis of adhesions was performed.The hernia sac was identified and dissected.The omentum was reduced.A ventralight st using echo ps (device #1) was placed and tacked.¿ (b)(6) 2017 - patient was diagnosed with small bowel obstruction and recurrent ventral hernia thereby underwent open repair with excision of ventralight st using echo ps (device #1) and implant of kugel hernia patch (device #2).Per op notes, ¿the prior mesh (device #1) was identified to be out of abdominal wall was excised.The adhesions of omentum to colon and small bowel were lysed.The obstruction was identified and released.The hernia defect was identified and excised.A kugel hernia patch (device #2) was placed using prolene sutures.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST W/ ECHO
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key12817865
Report Number1213643-2021-08732
Device Sequence Number20273816
Product Code FTL
UDI-Device Identifier00801741031786
UDI-Public(01)00801741031786(17)170128(10)HUZA0740
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K122436
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/28/2017
Device Catalogue Number5955810
Device Lot NumberHUZA0740
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/12/2021
Supplement Date Received by Manufacturer07/15/2025
Initial Report FDA Received Date11/15/2021
Supplement Report FDA Received Date08/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Disability;
Patient Age51 YR
Patient SexMale
-
-