• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SOLERA VOYAGER SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SOLERA VOYAGER SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 55750017545
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Pain (1994); Intraoperative Pain (2662)
Event Date 10/13/2021
Event Type  Injury  
Event Description
Information was received from a healthcare provider via a manufacturer representative regarding a patient with spinal therapy.It was reported that the set screw backed out.The l4 screw was also a little loose.As the back pain worsened, additional surgery performed to remove the screws and replace them.Since there was a possibility that the dorsal side of the left cage (capstonepeekptc) that had been placed was interfering with the nerve root, the posterior decompression of the left side was also performed additionally since the cage was sinking.There were no further complications reported regarding the event.The allegation of screw is looseness.There was no defect with the rod.
 
Manufacturer Narrative
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis :55750017545 lot# h5685263 after visual and optical examination and functional testing, it does not appear to be any damage, nor does it indicate any functional issues with the screw.D6a - implanted date updated d6b- explanted date updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SOLERA VOYAGER SPINAL SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12818125
MDR Text Key280808368
Report Number1030489-2021-01392
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00643169783911
UDI-Public00643169783911
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K170679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number55750017545
Device Catalogue Number55750017545
Device Lot NumberH5685263
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2021
Initial Date FDA Received11/16/2021
Supplement Dates Manufacturer Received03/30/2022
Supplement Dates FDA Received04/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age51 YR
Patient SexMale
-
-