| Catalog Number |
705.13P |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that tip of the forceps failed to open and close when inserting into the eye tissue during a cataract/vitrectomy combination procedure.The procedure was completed successfully after replacing the forceps with another one.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A sample is available that has not yet been received at manufacturing for evaluation.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A customer reported that opening and closing failure occurred to the tip of the forceps when using it during insertion.Sample received in opened original packaging including cover foil.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The received forceps sample was visually inspected with the aid of a photomicroscope with various magnifications.After cleaning, it was found that the tube is loose which block the forceps from activating.The customer¿s complaint was confirmed.Root cause was unable to verify.Why the tube was loose and blocked the forceps from activating could not be determined.This device passed the 100% control and was conform when leaving the production.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|