ARTHREX, INC. 35MM CENTRAL SCREW, MODULAR; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number 35MM CENTRAL SCREW, MODULAR
Health Effect - Clinical Codes Failure of Implant (1924); Foreign Body In Patient (2687)
Date of Event 10/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint confirmed, the ar-9560-24-4 and mating ar-9561-35s did not mate.Nothing broke off from the devices.The hex features were found to be damaged, indicating the devices were attempted to be assembled while they were not aligned.
 
Event or Problem Description
On 10/29/2021, it was reported by a sales representative via e-mail that an ar-9560-24-4 and ar-9561-35s baseplate broke off, leaving screw inside of the patient.The surgeon was able to remove the fragment.This was discovered during a procedure on (b)(6) 2021.The case was completed by using a new ar-9560-24-4 and ar-9561-35s.Additional information provided 10/29/2021: procedure being performed was a reverse total shoulder.The broken screw was removed from the patient.Bone quality of the patient was good.The 35mm tap was used to prepare the bone socket.The case was completed by using the 4.0 hex driver from mgs 2 to remove the screw.We opened a new baseplate and central screw (downsized to 30mm) and were able to successfully complete the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
35MM CENTRAL SCREW, MODULAR
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12819073
Report Number1220246-2021-03940
Device Sequence Number1792221
Product Code PHX
UDI-Device Identifier00888867296589
UDI-Public00888867296589
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model Number35MM CENTRAL SCREW, MODULAR
Device Catalogue NumberAR-9561-35S
Device Lot Number8765
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2021
Initial Report FDA Received Date11/16/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-