Catalog Number 0684-00-0513 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/22/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during insertion of the intra-aortic balloon (iab), the iab could not be passed through the patient's blood vessel.The iab was removed and replaced with a new one.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
Complete reporter name: (b)(6).Event site postal code: (b)(6).The device will not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).Device not returned.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period nov-2019 through oct-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #539172.
|
|
Event Description
|
N/a.
|
|