• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD TRANS-RAY 7FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD TRANS-RAY 7FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0513
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Event Description
It was reported that during insertion of the intra-aortic balloon (iab), the iab could not be passed through the patient's blood vessel.The iab was removed and replaced with a new one.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
Complete reporter name: (b)(6).Event site postal code: (b)(6).The device will not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).Device not returned.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period nov-2019 through oct-2021 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #539172.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANS-RAY 7FR. 34CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12819297
MDR Text Key280811040
Report Number2248146-2021-00750
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/14/2023
Device Catalogue Number0684-00-0513
Device Lot Number3000121125
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/14/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexFemale
Patient Weight62 KG
-
-