| Catalog Number |
03P76-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Respiratory Distress Syndrome of Newborns (2046)
|
| Date of Event |
10/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 27-oct-2021, abbott point of care was contacted by a customer regarding i-stat eg7+ cartridges that yielded a suspected descrepant potassium result of 6.1 mmol/l on an (b)(6) male with respiratory distress.There was no additional patient information available at the time of this report.Return product is not available for investigation.Method result (k): i- stat 6.1 mmol/l.Lab 3.9 mmol/l.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 15-nov-2021.A review of the device history record confirmed that cartridge lot n21173 passed finished goods release criteria.Retained cartridge test results met the acceptance criteria found in q04.01.003 rev.Ag, appendix 1- product complaint level 2 and level 3 investigation procedure.A previously identified deficiency was identified for eg7+ lot n21173 with respect to an elevated rate of suppressed results; punctures in the cartridge pouches were observed and are under investigation according to quality record (b)(4).
|
| |
|