| Model Number |
01-04-0020 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Hba1c control i of lot 12160 observed with values above the acceptable target range.Vendor identified an uncommonly high glucose level in the control material which glycates over time created false high values.Product lot had been placed on shipping hold near the same time as the complaint.No known reporting to medical staff or patients.
|
| |
|
Search Alerts/Recalls
|