| Model Number |
10471067 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A discordant falsely elevated high-sensitivity troponin i (tnih) patient sample result was obtained on a dimension vista 500 system.The discordant result was reported to the physician(s).The same sample was reprocessed after removal of correlation factors for the tnih assay.A lower result was obtained.There are no known reports of patient intervention or adverse health consequences due to the discordant falsely elevated patient tnih result.
|
| |
|
Additional Manufacturer Narrative
|
|
A united states customer contacted the siemens healthcare diagnostics customer care center (ccc) regarding a discordant falsely elevated high-sensitivity troponin i (tnih) patient sample result obtained on a dimension vista 500 system.Siemens is investigating the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information (21-nov-2021): siemens healthcare diagnostics headquarters support center (hsc) concluded the investigation of the event.Hsc reviewed the instrument data log and the information provided by the customer.Hsc concluded that the event was an operator use error.The dimension vista high sensitivity troponin i (tnih) assay lot 21209ba does not require the use of lot specific correlation factors.The cause of the event was that the customer was operating with tnih lot specific correlation factors entered for a tnih assay lot that did not require lot specific correlation factors.Removal of the incorrect tnih assay correlation factors resolved the issue.The device is performing within specifications.No further evaluation is required.Section h6 has been updated to reflect the hsc investigation initial mdr 2517506-2021-00308 was filed 16-nov-2021.
|
| |
|
Search Alerts/Recalls
|