• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CERAMAX CERAMIC INSERT 36 X 52; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CERAMAX CERAMIC INSERT 36 X 52; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS Back to Search Results
Model Number 1218-87-652
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 10/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the surgeon said the patient complained of pain in her hip.He believed the cup may be loose, and wanted to remove it and replace with a revision cup.After removing the liner and screws he inspected the cup and found that the cup was not loose.He implanted new screws, head and a constrained liner.He was unable to get the locking ring to seat on the constrained liner due to visualization and decided to remove it.He replaced it with a new +4, 10 degree liner and appropriate size head.Doi: (b)(6) 2018, dor: (b)(6) 2021, affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot for allegations of pain a device manufacturing (mre) review will not be performed even when product/lot information is known.Per wi-3430, review of the device history record is unlikely to add value to the complaint investigation regarding an allegation of pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERAMAX CERAMIC INSERT 36 X 52
Common Device Name
PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12820648
Report Number1818910-2021-25195
Device Sequence Number4906206
Product Code MRA
UDI-Device Identifier10603295012511
UDI-Public10603295012511
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1218-87-652
Device Catalogue Number121887652
Device Lot Number8586444
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2021
Supplement Date Received by Manufacturer12/07/2021
Initial Report FDA Received Date11/16/2021
Supplement Report FDA Received Date12/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/25/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CERAMAX CERAMIC INSERT 36 X 52; DLT TS CER HD 12/14 32MM +5; DLT TS CER HD 12/14 36MM +8.5; PINN CAN BONE SCREW 6.5MMX20MM; PINN CAN BONE SCREW 6.5MMX25MM; PINN LNR CON +4 10D 32IDX52OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-