Complainant part is expected to be returned for manufacturer review/ investigation but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on (b)(6) 2021, during a proximal humerus, the surgeon had difficulties getting the wire guide sleeve for 3.7 cannulated screws to thread into the plate.Ultimately, he switched to using a regular locking to we, a wire sleeve to measure the screw and placed it successfully.Procedure was completed successfully with five(5) minutes surgical delay.No patient consequences.This report is for one (1) 1.6 wire guide for 3.7 cann lckng screws.This is report 1 of 2 for complaint (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: updated data: d4, h4.Investigation summary visual inspection: the 1.6 wire guide for 3.7 cann lckng screws (part# 360.266, lot# 6352482) was returned and received at us cq.Upon visual inspection of the complaint device no issues were observed.Functional test: a functional assessment was not performed on the complaint device since it was returned without relevant mating device.Can the complaint be replicated with the returned device(s)? unable to perform.Document/specification review: based on the date of manufacture, the current and manufactured revision of drawings were reviewed no design issues or discrepancies were identified.Complaint confirmed? no.Investigation conclusion: this complaint was not confirmed since no issue was observed with device.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot part # 360.266.Synthes lot # 6175616.Supplier lot # na.Release to warehouse date: 14 jul 2009.Manufactured by synthes elmira.No ncr's were generated during production.Device history review: review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Corrected data: d9, g1, h3, h6.
|