| Model Number |
24690 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the moderately tortuous and calcified internal carotid artery.A 3.0mmx30mmx135cm sterling balloon catheter was advanced for post-dilation.However, during inflation, the balloon ruptured.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the moderately tortuous and calcified internal carotid artery.A 3.0mmx30mmx135cm sterling balloon catheter was advanced for post-dilation.However, during inflation, the balloon ruptured.The procedure was completed with a different device.There were no patient complications nor injuries reported.It was further reported that the balloon ruptured while increasing the pressure from 3 atmospheres to 4 atmospheres during the initial inflation.
|
| |
|