| Model Number |
68LP-420 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/27/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.
|
| |
|
Event or Problem Description
|
|
Device explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: opening, wear abrasion, crease fold.Leak test was performed and found opening.A microscopic analysis was performed which identify crease smooth opening on posterior.The fill test inspection was performed, the result is no blockage based on the device analysis the final assessment is: -a crease smooth opening on posterior assessed as fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device explanted.
|
| |
|