|
Catalog Number OM-7500 |
Device Problems
Material Twisted/Bent (2981); Firing Problem (4011)
|
Patient Problems
Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/28/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference: case- (b)(4).
|
|
Event Description
|
It was reported that during a shoulder arthroscopy procedure, there was a failure of the speedlock implant.The sutures were not pulled into the shaft of the inserter into the implant.There was excessive suture into the joint and there was no salvage.The procedure was successfully completed with non-significant delay using a s+n back-up device.No further complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.A visual evaluation of the device showed it was not received in any original packaging.The wire snare ring and the anchor were not returned.The long snare wire has been coiled up on the suture reel.The distal end of the insertion device is bent.A functional test of device one is not possible since the device is intended for single use.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states based on the information provided, the implantable sutures was retained in the patient¿s joint it was left secure.Furthermore, micro-motion and/or migration of the retained sutures is unlikely.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include issues encountered while loading the suture into the device or introduction into the cavity.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods.No containment or corrective actions are recommended at this time.
|
|
Manufacturer Narrative
|
H10¨ h3, h6¨the reported device was received for evaluation.A visual evaluation of the device showed it was not received in any original packaging.The wire snare ring and the anchor were not returned.The long snare wire has been coiled up on the suture reel.The distal end of the insertion device is bent.A functional test of device one is not possible since the device is intended for single use.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states based on the information provided, a user error was the likely cause of the reported adverse event.Since the implantable sutures was retained in the patient¿s joint it was left secure.Furthermore, micro-motion and/or migration of the retained sutures is unlikely.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force, do not bend or twist the inserter handle during and after insertion as damage to the implant or incomplete insertion may result; the system requires tension to be distributed through the suture ends to achieve suture lock.No containment or corrective actions are recommended at this time.
|
|
Manufacturer Narrative
|
H10: internal complaint reference case (b)(4).H3, h6: the reported device was received for evaluation.A visual evaluation of the device showed it was not received in any original packaging.The wire snare ring and the anchor were not returned.The long snare wire has been coiled up on the suture reel.The distal end of the insertion device is bent.A functional test of device one is not possible since the device is intended for single use.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states based on the information provided, the implantable sutures was retained in the patient¿s joint it was left secure.Furthermore, micro-motion and/or migration of the retained sutures is unlikely.The root cause has been associated with the use of the device.Factors that could have contributed to the reported event include issues encountered while loading the suture into the device or introduction into the cavity.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|