| Model Number |
FT4 G2 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received questionable elecsys ft4 ii results for one patient tested on a cobas 6000 e601 module with serial number (b)(4).The initial result was reported to the clinician, who asked that the sample be rerun on an abbott instrument.The patient's sample had an initial result of 2.24 ng/dl.On (b)(6) 2021, the repeat result was 0.88 ng/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
The alarm trace, pre-analytical and calibration data did not indicate an issue.The investigation did not identify a general reagent problem.The cause of the event could not be determined.
|
| |
|