| Model Number |
US906 |
| Medical Device Problem Code |
Unexpected Color (4055)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer evaluation: investigation on-going.Should relevant additional / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap inc.That a steam ster locks orange (part # us906) was used to secure a sterilization container.According to the complainant, the indicator dot failed to consistently transition to its post sterilization color.Reportedly, one side of the lock would change color, but the blue mark on the other side would not completely transition.Color would not even partially change from blue to black.The set was pulled, the lock was changed, and the instruments were re-sterilized.The complaint device has been returned to the manufacturer for evaluation.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer evaluation: the device was returned to the manufacturer for physical evaluation.Functional testing was performed by exposing twelve (12) of the locks to steam cycles; a visual examination of the chemical indicator revealed that all dots transitioned from blue to brown.The device history records (dhr) were reviewed for all available lot numbers; the devices were found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.The investigation into the cause of the reported problem was not able to confirm the failure mode of chemical indicator failed to transition to its post sterilization color.
|
| |
|
Search Alerts/Recalls
|