Model Number 2307-72-003
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Instrument deltopectoral cutting guide faulty.Grub screw in the square hole is cross threaded/stiff/stuck, needs to be replaced.Surgical delay of 2 minutes.Procedure completed successfully.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned for analysis.All available photos were reviewed and the failure mode can be confirmed.Root cause attributed to the device being worn from normal use and servicing.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTO PECTORAL CUTTING GUIDE
Common Device Name
EXTREMITY INSTRUMENTS : RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY INT'L LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK  LS11 8 DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12822033
Report Number1818910-2021-25237
Device Sequence Number5696406
Product Code HTZ
UDI-Device Identifier10603295116080
UDI-Public10603295116080
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2307-72-003
Device Catalogue Number230772003
Device Lot Number2382469
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2021
Supplement Date Received by Manufacturer01/12/2022
Initial Report FDA Received Date11/16/2021
Supplement Report FDA Received Date01/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/26/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-