• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD UNKNOWN INTRA-AORTIC BALLOON; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD UNKNOWN INTRA-AORTIC BALLOON; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number UNKNOWN
Device Problems Optical Problem (3001); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2021
Event Type  malfunction  
Manufacturer Narrative
Occupation: sicu manager.(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.(b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab), the customer had questions regarding the arterial waveform and numbers.They then stated that the fiber optic cable hadn¿t been working.Troubleshooting revealed a prior fiber optic sensor failure.Balloon tip position was confirmed by the cardiologist.All cables and connections were verified to be secure.The customer aspirated the line getting back adequate blood return.The arterial waveform displayed with better augmentation.Maintenance practices and leveling were reviewed.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint #(b)(4).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN INTRA-AORTIC BALLOON
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12823694
MDR Text Key282566514
Report Number2248146-2021-00757
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/26/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight100 KG
-
-