| Model Number |
101-85555F |
| Medical Device Problem Codes |
Unintended Movement (3026); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the denali rod connector was discovered to be loose during a revision procedure.The procedure was completed successfully.No adverse consequences nor medical intervention were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device and complaint history records were reviewed for this lot, no relevant manufacturing issues or similar complaints were identified.No additional information was provided about the event.Dates of the initial and re-spondylodesis surgeries and unknown.Post-op activity and fusion level is unknown.The root cause of the reported event cannot be determined concussively from the information provided and without device evaluation.Potential causes include: - excessive patient post-op activities beyond surgeon recommendation.- patient non-fusion.- poor fixation construct.- device used beyond intended us.
|
| |
|
Event or Problem Description
|
|
It was reported that the denali rod connector was discovered to be loose during a revision procedure.The procedure was completed successfully.No adverse consequences nor medical intervention were reported.
|
| |
|