Model Number VNL8-J10 |
Device Problems
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/11/2021 |
Event Type
malfunction
|
Event Description
|
Noise and blackout occur when the angle is applied during use.This happens especially when you put up.No trauma can be confirmed at the insertion part, and there is no leak.There is no health hazard to patients, medical staff, etc.Due to the above failure situation.
|
|
Manufacturer Narrative
|
If additional information becomes available, a supplemental report will be filed with the new information.
|
|
Manufacturer Narrative
|
Evaluation summary: we checked the returned unit and confirmed that the h1 signal in the flexible printed board of the ccd module as conduction failure.Based on the result, we concluded that it was caused due to too more usage of repeated angulation than estimated lifetime.Therefore, h1 signal in the flexible printed board got damaged by the tension of ccd cable ,when the endoscope angulation function was used.Correction information: g6: follow up #1.H6: coding changed based on the investigation result.
|
|