| Model Number |
NAV2020K |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a healthcare professional (hcp) via a medtronic representative regarding patient was implanted with a screw using a driver in an unknown spinal therapy for an l3-s1 revision case.It was reported that the screwdriver tip broke when the screw was put halfway-in.The tip is still stuck inside the screw head under the saddle of the tulip.The screw was removed and changed to a new one, along with change of navigated screwdriver.There was no allegations aimed towards the screw.The bone was very hard according to the surgeon.The date of implant and explant for screw was (b)(6) 2021.There were no patient symptoms or complications as a result of the event.There was no treatment or additional surgery performed as a result of this event.No further complications were reported/ anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis- part# nav2020k ; lot# ca16g057- the portion of the driver head that interfaces with the screw has been broken/sheared off.The face of the break is relatively flat with a slight angle.This is consistent with a drive tip that has been sheared off from overload.H6: updated eval.Code method and eval.Code conclusion post analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|