• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5530-G-609-E
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2021
Type of Reportable Event Malfunction
Event or Problem Description
During poly impaction surgeon could not lock the poly triathlon mechanism into the tibia.The poly was removed and inspected and impacted again.At second attempt poly appeared seated and locked.Post op x-ray showed issue with locking mechanism.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned.
 
Event or Problem Description
During poly impaction surgeon could not lock the poly triathlon mechanism into the tibia.The poly was removed and inspected and impacted again.At second attempt poly appeared seated and locked.Post op x-ray showed issue with locking mechanism.
 
Additional Manufacturer Narrative
Reported event an event regarding seating/locking issues involving a triathlon insert was reported.The event was confirmed through clinician review of the provided x-rays.Method & results product evaluation and results: not performed as no product was returned for evaluation.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: on the ap x-rays of the right tka the locking wire of the tibial insert is seen to be displaced from its native position.The failure to properly seat the tibial insert with disruption of the locking mechanism is confirmed.The root cause of this failure can not be determined from this limited information.Should further documentation become available i would be happy to further this assessment.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the event reported during poly impaction surgeon could not lock the poly triathlon mechanism into the tibia.A review of the provided medical records by a clinical consultant stated the following comment: on the ap x-rays of the right tka the locking wire of the tibial insert is seen to be displaced from its native position.The failure to properly seat the tibial insert with disruption of the locking mechanism is confirmed.The root cause of this failure can not be determined from this limited information further information such as video/photographs of the event as it occurred as well as return of the device and primary operative reports are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12827842
Report Number0002249697-2021-01920
Device Sequence Number13823398
Product Code MBH
UDI-Device Identifier07613327383133
UDI-Public07613327383133
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172634
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model Number5530-G-609-E
Device Catalogue Number5530-G-609-E
Device Lot NumberRA4PTK
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2021
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date11/17/2021
Supplement Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age74 YR
Patient SexMale
-
-