• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 24 IN. TOUCH SCREEN; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 24 IN. TOUCH SCREEN; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10003
Medical Device Problem Codes No Display/Image (1183); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during set up for a cori assisted tka surgery, the real intelligence 24 in.Touch screen blacked out after registering hand piece.The words ¿no signal¿ came up on the screen and neither the touch screen nor the real intelligence tablet had any signal.The light above the tablet cord port was blinking.They rebooted the real intelligence cori and they could proceed.The procedure was completed with the same devices without significant delays.Since incident occurred before procedure, patient was not involved.
 
Additional Manufacturer Narrative
Internal reference number: case- (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the cori touchscreen pn rob10003,sn (b)(6), intended for used in treatment, was not returned for evaluation.A visual and functional evaluation could not be performed and a relationship between the reported event and the device could not be determined.A review of manufacturing records indicates the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.No reasonable contributing factors could be identified based on the received complaint information and investigation results.The product was not returned and no evidence was made available to link the complaint to an escalation event.In the event of a system failure, refer to the recovery procedure guidelines in the real intelligence cori for knee arthroplasty user manual.A failure can consist of, but is not limited to, a system software crash, unrecoverable hardware failure, handpiece failure with no backup available, tracker failure or loss of contact with bone that is unrecoverable, etc.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Smith & nephew has not received adequate materials to fully evaluate the complaint, but if additional relevant materials are later received the complaint can be reopened.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE 24 IN. TOUCH SCREEN
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
austin, TX 78735
5123913905
MDR Report Key12827919
Report Number3010266064-2021-00768
Device Sequence Number16488290
Product Code OLO
UDI-Device Identifier00885556756256
UDI-Public00885556756256
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10003
Device Catalogue NumberROB10003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2021
Supplement Date Received by Manufacturer02/21/2022
Initial Report FDA Received Date11/17/2021
Supplement Report FDA Received Date02/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-