| Model Number |
SCF-770 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade iv.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified, weight to the spec, crease flat, deformation, and device appearance (observed split in patch area, it is out of specification per qa (b)(4) was identified which is characterized as a workmanship, however this workmanship is not relationship with the complaint).After autoclave cycle were observed: cloudy color in the gel and voids.Based on the device analysis the final assessment is: split in patch area, assessed as a workmanship.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade iv.Device has been explanted.
|
| |
|