| Model Number |
227F3 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the 227f3 electrophysiology (ep) catheter with lot number 45117 was returned and analyzed.Visual inspection of the ep catheter showed the shaft was kinked at strain relief.Smart chip verification indicated the catheter was used for 1 injection.The catheter failed the performance test due to the system notice 50030 indicating "excessive flow".The pressure test revealed a leak through the shaft at strain relief.In conclusion, the ep catheter failed the returned product inspection due to the shaft breach.
|
| |
|
Event or Problem Description
|
|
After a completed case, the electrophysiology (ep) catheter tested out of specification per the manufacturer's investigation.
|
| |
|