| Catalog Number |
0026880 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject product was returned for evaluation and confirmed the reported condition of a missing component.Based on the condition of the sample this is identified as an out of the box condition.The customer reported the cap was not within the sterile package.Based on the sample evaluation the complaint is confirmed as a manufacturing error.Review of the device history records identified no anomalies in production of this lot of (b)(4) units.Release in june 2021.No other complaints have been reported to date for this lot.
|
| |
|
Event or Problem Description
|
|
As reported, during a laparoscopic cholecystectomy procedure on (b)(6) 2021, the bard/davol hydrosurg plus laparoscopic irrigator device was presented for use.It was identified that the threaded cap on the nd handle was missing.The product packaging is reported as being completely sealed prior to opening for the case and noting the missing component.As reported, there was no prolongation of surgery time and another device was used to complete the procedure.There was no reported patient injury.Attempted use of the device with the cap missing would result in fluid leak.
|
| |
|