U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SPINE BLOCKER; INSTINCT JAVA SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SPINE BLOCKER; INSTINCT JAVA SYSTEM Back to Search Results
Catalog Number 046W0AN00002
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that five blockers were stripped during surgery during final tightening.The surgeon had trouble locking them, so removed them an found the damage.This is report four of five for this event.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3003853072-2021-00124 through 3003853072-2021-00128.
 
Event or Problem Description
It was reported that five blockers were stripped during surgery during final tightening.The surgeon had trouble locking them, so removed them an found the damage.This is report four of five for this event.
 
Additional Manufacturer Narrative
Device evaluation: visual inspection revealed that the screw drive of each blocker was stripped.Potential cause: root cause was unable to be determined.This event could possibly be attributed to off-axis forces applied during plug insertion.Dhr review: per dhr review, the part was likely conforming when it left zimmer biomet control.Device use: this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOCKER
Common Device Name
INSTINCT JAVA SYSTEM
Manufacturer (Section D)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR  33080
Manufacturer (Section G)
ZIMMER SPINE
23 parvis des chartrons
cite mondiale
bordeaux, cedex 33080
FR   33080
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key12830878
Report Number3003853072-2021-00127
Device Sequence Number17234468
Product Code NKB
UDI-Device Identifier00889024339910
UDI-Public(01)00889024339910(10)V20181(11)210315
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K111301
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number046W0AN00002
Device Lot NumberV20181
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2021
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date11/17/2021
Supplement Report FDA Received Date03/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexMale
Patient Weight77 KG
-
-