• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION SIMULUS SEMI-RIGID ANNULOPASTY BAND; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION SIMULUS SEMI-RIGID ANNULOPASTY BAND; RING, ANNULOPLASTY Back to Search Results
Model Number 800SC24
Medical Device Problem Codes Material Separation (1562); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Endocarditis (1834); Pseudoaneurysm (2605); Mitral Valve Insufficiency/ Regurgitation (4451); Insufficient Information (4580)
Date of Event 10/21/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that 1 year and 4 months post implant of this 24mm mitral annuloplasty band, it was explanted and replaced with a 25mm mitral bioprosthetic valve. the reason for the replacement was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic received additional information that changed the event description and device relatedness of this previously reported event.Additional information was received that this mitral annuloplasty band was implanted originally due to mitral regurgitation from endocarditis.Then approximately 1 year and 4 months later, the patient presented with repeat methicillin resistant staph aureus endocarditis.This was associated with a sub-annular abscess, which resulted in a ventricular septal defect (vsd), recurrent severe mitral valve regurgitation and a pseudoaneurysm in the septal wall.During re-operation, it was noted that the posterior annulus of the annuloplasty band was dehisced.The patient¿s annulus was repaired, mitral valve replaced, and vsd was closed.The surgeon reported that there was no defect with the annuloplasty device.No additional adverse patient effects were reported.There is no evidence to suggest the device caused or contributed to a death or serious injury.Valid scientific evidence supports no valve-related cause for endocarditis where implant time exceeds two months.Ref: interventional and surgical cardiovascular pathology, isbn 0-7216-2457-x, p.38 and p.142.Updated: a4, b3, b7 and h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMULUS SEMI-RIGID ANNULOPASTY BAND
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key12831544
Report Number2025587-2021-03473
Device Sequence Number14559336
Product Code KRH
UDI-Device Identifier00643169182202
UDI-Public00643169182202
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K083683
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model Number800SC24
Device Catalogue Number800SC24
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2021
Supplement Date Received by Manufacturer12/13/2021
Initial Report FDA Received Date11/17/2021
Supplement Report FDA Received Date01/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age22 YR
Patient SexMale
Patient Weight119 KG
-
-