Model Number 71362292
Medical Device Problem Codes Break (1069); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal reference number: (b)(4).
 
Event or Problem Description
It was reported that a u-joint screwdriver came through the wash broken.Articulating head has seized up.As this was noticed upon cleaning and sterilization activities, there was no patient involvement.
 
Additional Manufacturer Narrative
Lot number and expiration date added.Manufacture date added.(b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was returned for evaluation.A visual inspection of the returned u-joint screwdriver confirms the head is seized up.The device shows signs of significant wear and use.The contribution of the device to the reported event could not be corroborated.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this product and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
U-JOINT SCREWDRIVER
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
austin, TX 78735
5123913905
MDR Report Key12831828
Report Number1020279-2021-08231
Device Sequence Number11905036
Product Code HRY
UDI-Device Identifier03596010468253
UDI-Public03596010468253
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71362292
Device Catalogue Number71362292
Device Lot Number17MM07917
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2021
Supplement Date Received by Manufacturer11/30/2021
12/28/2021
Initial Report FDA Received Date11/17/2021
Supplement Report FDA Received Date12/01/2021
12/28/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/12/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-