| Model Number |
TMTWB11 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/06/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).Patient age is not provided / unknown.Initial reporter¿s title is not provided / unknown.Patient weight is not provided / unknown.
|
| |
|
Event or Problem Description
|
|
Dr.Indicated lack of primary stability.Implant lost its primary stability completely during the placement.Tooth location # ul6 (palmer).
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.The following sections are being reported: dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number for similar event and no other complaint was identified.As documented in the summary investigation, contributing factors for this reported event likely exist outside of zimmer biomet control, including those related to patient biological factors/condition and surgical technique.Dental implants are designed and manufactured to achieve osseointegration after placement into the osteotomy using the recommended surgical protocol.The inability to achieve primary stability is a well-documented, previously investigated failure which is currently trended on a monthly basis.Current implant design and manufacturing risk documentation assess unable to place implant into osteotomy as potential failures, ultimately through osseointegration failure, with adequate controls in place.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for the inability to place implants into osteotomy have not identified any signals indicating potential manufacturing non-conformances impacting primary stability.With no signals indicating a systemic quality issue, no escalation to capa is required.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|