Model Number TVC-MC10
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2021
Type of Reportable Event Malfunction
Event or Problem Description
One of the front wheel casters of the mc10 imaging system was reported to have almost fallen off of the console while it was being moved.No patient involvement was reported.
 
Additional Manufacturer Narrative
Investigation of the product problem concluded that due to the design of the cart, a caster coming off the console did not present a risk of tipping/mechanical hazard.Moreover, caster screws on carts built since 2020 are subject to application of thread-locking adhesive and a torque specification.The cart involved in this complaint was built in 2016 prior to the use of thread-locking adhesive and a torque specification.In response to this complaint, the tvc-mc10 field service manual was revised to include a caster bolt check at preventative maintenance.In addition, the design fmea was updated to add an additional attribute specification regarding console casters.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKOTO INTRAVASCULAR IMAGING SYSTEM¿
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
INFRAREDX, INC.
28 crosby drive, suite 100
bedford MA 01730
Manufacturer (Section G)
INFRAREDX, INC.
28 crosby drive, suite 100
bedford MA 01730
Manufacturer Contact
stephen sum
28 crosby drive, suite 100
bedford, MA 01730
7813459651
MDR Report Key12832643
Report Number3015551113-2021-00001
Device Sequence Number4316376
Product Code IYO
UDI-Device Identifier00857595006432
UDI-Public(01)00857595006432(11)180509
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K163345
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTVC-MC10
Device Catalogue NumberTVC-MC10
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2021
Supplement Date Received by Manufacturer11/12/2021
Initial Report FDA Received Date11/17/2021
Supplement Report FDA Received Date02/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/09/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-