The manufacturer previously received information alleging an issue related to a remstar auto a-flex device's sound abatement foam.The patient alleged to have seen microfibers in the mask and tubing and experiencing loss of breath.There was no report of serious patient harm or injury.There was no report of medical intervention.The device was returned to a third-party service center for further evaluation.During evaluation, sound abatement foam degradation was not observed.There was no evidence of foam degradation or particles in the device.The device did not have any type of contamination.The device was found to be functional.The service center retrieved and reviewed the device's error logs.There were 6 errors found.The third-party service center could not confirm the customer's allegation and there was no visible foam degradation.The unit was scrapped after device evaluation was completed.Sections g1, h2, h3, and h6 have been updated in this report.
|