Model Number G3603522
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/26/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from healthcare professional via field representative regarding patient with pre-operative diagnosis- c1/2 subluxation.Involved in the c1/2 fusion procedure.It was reported that intra-operatively, after placement, when compression was applied, the position of the rod changed from the screw head, the connector was changed.During screw insertion, there was bleeding (probably the venous plexus) and the screw insertion was stopped.The screw was changed to a shorter screw that does not reach the va when reinsertion.Product was used correctly according to the directions given in the ifu/labeling.There was delay in overall procedure time for less than 60 minutes.Patient was not hospitalized prolong.On 2021-nov-04, received additional information that hemostasis was performed with floseal and gauze.
 
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.(b)(4) used for hemostasis performed in the patient.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluated summary- after visual and optical examination and functional testing, it does not appear to be any damage, nor does it indicate any functional issues with the screw.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ OCCIPITOCERVICAL UPPER THORACIC SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12834485
Report Number1030489-2021-01431
Device Sequence Number19965591
Product Code KWP
UDI-Device Identifier00643169542280
UDI-Public00643169542280
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K163375
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG3603522
Device Catalogue NumberG3603522
Device Lot NumberH5612016
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2021
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date11/18/2021
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age79 YR
Patient SexMale
-
-