• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problems Optical Problem (3001); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate this unit.Upon inspection by the fse, no failure could be reproduced.This unit passed all functional and safety tests per factory specifications.The iabp unit has been returned to the customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during patient use the cardiosave intra-aortic balloon pump (iabp) had a fiber optic module failure.There was no harm or injury to the patient and no adverse event was reported.
 
Manufacturer Narrative
Updated fields: b4, b5, d11, g4, g7, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: h6 (problem code).
 
Event Description
It was reported that during patient use the cardiosave intra-aortic balloon pump (iabp) had a fiber optic module failure.It was later clarified by the customer that it was an iab sensor failure.There was no harm or injury to the patient and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
brian schaeffer
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12835749
MDR Text Key282722034
Report Number2249723-2021-02669
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2021
Initial Date FDA Received11/18/2021
Supplement Dates Manufacturer Received06/15/2022
Supplement Dates FDA Received06/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/03/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
INTRA-AORTIC BALLOON CATHETER
-
-